English-28STAND-ALONE CALIBRATIONStand alone calibration can recalibrate the display without using a computer by the integrated front sensor or connecting anexternal sensor. There are 5 functions in STAND-ALONE CALIBRATION.• Self Calibration: Calibrates the integrated front sensor based on the USB color sensor. Alternatively, manually adjust theintegrated front sensor with an external near range color analyzer reference, which is traceable to a primary standardslaboratory.• Copy Calibration: Copies the luminance from one display to one or more additional displays.• Gamma Adjust: Recalibrates Gamma data by the integrated front sensor.• DICOM Measurement: Measures DICOM response.• AMBIENT SENSOR CALIBRATION: To manually adjust the integrated ambient light sensor with an external color analyzerreference, which is traceable to a primary standards laboratory.NOTE: Stand before proper Stand-Alone calibration can be performed, the display should warm-up for a minimum of30 minutes. If Stand-Alone calibration is started before the monitor is warmed up, a warning will appear on-screen(Figure A.1).NOTE: Please only use an USB color sensor according to our specifi cation.Refer to the KEY MAP (Figure A.2) to adjust the settings for calibration.Stand-Alone calibration should be performed in either portrait or landscape orientation.NOTE: To start calibration without a computer, perform the following steps.When the “NO SIGNAL” message appears on screen, press the “LEFT” and then “RIGHT” buttons and hold themdown simultaneously. The CALIBRATION menu will appear.Figure A.2KEY MAPUP/DOWN: Changes from setting to settingLEFT/RIGHT: Changes the setting selection(ie...SELF or COPY)SELECT: Moves to the next step of calibrationEXIT: Moves back one step of calibrationFigure A.1CAUTION: Please use a properly-managed USB color sensor. If there is any doubt about the result of the Stand-Alone calibration result, please contact your supplier.