English-30Copy CalibrationThe white point and luminance can be copied from one display to one or more additional displays. Using this feature reducesthe variation between different displays allowing them to match each other more closely.ABUSB color sensorFigure C.1Display A – SOURCE display of white point and luminance to copy.Display B – MD210C2/MD211C2/MD211C3 which performs a copy.ABFigure C.2Sensor Port(located at the back of the monitor)MD210C2/MD211C2/MD211C3MD210C2/MD211C2/MD211C3Source display Sourcedisplay1. To show the CALIBRATION menu, plug-in the USB color sensor to the sensor port (Figure C.1) or select the STAND ALONECALIBRATION in the advanced menu (page 22).2. Set the target white color on Display A. For example, enter the Advanced OSD by pressing RESET and EXIT.3. Using “LEFT” or “RIGHT” of Display B, select COPY in the MODE selection (Figure C.3).4. The procedure will ask for the USB color sensor to be placed on the center of the display panel (Figure C.4). Tilt the displaypanel approximately 5° backward and place the USB color sensor in the center of the display panel (Figure C.1).NOTE: Place the USB color sensor fl at against the display A to avoid any infl uence of ambient light. Do not press the USBcolor sensor against the display panel. Press “SELECT”.NOTE: Make sure that the USB color sensor cable is positioned to avoid blocking the ambient light and human sensors.5. Press SELECT on Display B to start measuring the white point and luminance of Display A.6. After copying Display A information, the target luminance will be stored and displayed in the OSD of Display B. The targetvalue is not adjustable (Figure C.5). The white points of Display A and Display B should match.Press SELECT if the copy result is satisfactory. If the copy result is unsatisfactory, press “RESET”.7. When you want to confi rm the white point of Display B, remove the USB color sensor from source Display A and place inthe center of Display B (Figure C.2). To skip the white point confi rmation, Display B, press “SELECT” to step 9 for FINETUNING.8. Press SELECT to start the confi rmation of white point.9. After the confi rmation is fi nished, the FINE TUNING MODE message appears.10. At the FINE TUNING MODE message (Figure C.6), press SELECT if the copy result is satisfactory. If the copy result isunsatisfactory, manually fi ne tune the white point using the “LEFT” and “RIGHT”, and press SELECT.11. After completing the calibration, you can set DATE and TIME using control keys.Figure C.4 Figure C.5Figure C.3 Figure C.6