Defender 5000 Indicators EN-414.3.3 Application SetupThe Application can be customized for various user preferences.Press the Softkey corresponding to the iconto enter Configuration Setup.The Configuration Menu is now displayed.Select the list item and press the Softkeycorresponding to the icon ,to change thesetting as desired.To return to the Application home screen, pressthe button corresponding to the icon .The Check Configurations are defined below (defaults in Bold).Configure Item Option (Bold is default) DescriptionCheck Mode Weighting/Counting Weighing: Check weighing mode.Counting: Check counting mode.Auto Tare Off/On Off: Auto tare is turned off.On: The first stable weight (>=5d) will be tared as acontainer automatically.Accumulate Off/Manual/Automatic Off: The Softkey "∑" does not display.Manual: The Softkey "∑" displays. User can press the keyto accumulate the stable weight.Automatic: The Softkey "∑" displays. The weight will beaccumulated automatically.Note:The load to be accumulated has to be >= 5d and the nextaccumulation can only start once the pan has been cleared.When LFT is ON(no such limit when LFT is OFF or theapprove mod is OIML),a. gross weight and net weight cannot be accumulated;b. gross 0 has to be reached before a new sampleaccumulation.When LFT is OFF,a. gross weight and net weight can be accumulated;b. a new sample can be accumulated after reaching gross0 or net 0.Audible Signal Off/Under andOver/Accept/Under/OverOff: No beep.Under and Over/Accept/Under/Over: Beep whenreaching the selected check point.Auto Opt. Off/On Off: Auto Opt. is off.On: The APW will be optimized automatically when doingcount weighing.Opt. Beep Off/On Off: Opt. Beep is off.On: When the APW is optimized, the beeper will beeponce.APW Auto Save Off/On Off: APW Auto Save is off.On: If the APW is get from sampling, and a countinglibrary is selected, the new APW will save to library afteroptimized.Note: It will be hidden when "Auto Opt." is off.