Defender 5000 Indicators EN-47Configure Item Option(Bold is default) DescriptionNote:1. The load to be accumulated has to be >= 5d.Another accumulation can't be done until thepan is cleared (< 5d).2. Gross weight and net weight can't beaccumulated together when the LFT is ON (nosuch limit when the LFT is OFF or theapproved model is OIML). When firstaccumulated value is gross weight (netweight), scale will enter into gross weight (netweight) accumulate mode. Otherwise, “Grossand net weight can't be accumulated” will bedisplayed.3. If the LFT is ON, gross weight 0 has to bereached before a new sample can beaccumulated. If the LFT is OFF, new samplecan be accumulated after reaching grossweight 0 or net weight 0.Duration Time 1 ~ 10 s It is the time for the display to remain the dynamicweighing result after the load is removed.4.5.2 Average Time SetupThere are two methods to set the averaging time.The default Average Time is 10 seconds.Note: When the time is set to 0, the first stable weight over 5d will be displayed.Averaging time can be set to a value between 0 and 60 seconds.Method 11. Press the button corresponding to the iconto enter the Dynamic sub-menu forchanging the averaging time.2. Press the Softkey corresponding to the iconand then key in the desired value using thealphanumerical keypad.3. After setting the averaging value desired,press the Softkey corresponding to the iconfor confirmation.4. Press the Softkey corresponding to the iconto exit the sub-menu.