19 Defender 5000 Indicators Service Manual>|1|Specify the graduation for weighing range 1. If Single range is enabled, this will be the graduation for theentire weighing range of the scale. If Dual range is enabled, this will be the graduation used in the lowerrange.>|2|Specify the graduation for range 2. If Dual range is enabled, this will be the graduation for the secondweighing range of the scale. If Single range is enabled, this parameter will not be shown.NOTE: Graduation settings are limited to values from Capacity divided by 600 to Capacity divided by75000. Therefore, not all settings are available for each capacity.2.4.5 LanguageSet the language displayed for menus and displayed messages.EnglishDeutschFrançaisItalianoPolskiSpanish한국中文日本语2.4.6 Power On ZeroZero the balance at Power On.Off = disabled.On = enabled.2.4.7 Power On UnitSet the unit that will be displayed at Power On.Automaticgkglbozlb:ozt (Metric Tonne)ton (Short Ton)2.4.8 Key BeepSet how the beeper sounds when a key is pressed.Off = no soundOn = sound2.4.9 Transaction CounterThe transaction counter is a seven-digit counter that tracks the total transactions. When the value reaches9,999,999, the next transaction causes a roll-over to 0000001.Off = the transaction counter will not increase.On = the transaction counter will increase with the additional menu item Next Transaction available.