24 Defender 5000 Indicators Service Manual2.7.8 Custom Unit (c)Use the Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using aconversion factor, where the conversion factor is the number of custom units per gram expressed inscientific notation (Factor x 10^Exponent).FactorSet the conversion factor using the numeric keypad.Settings of 0.1000000 to 1.9999999 are available. The default setting is 1.0.ExponentSet the factor multiplier.-3 = divide the Factor by 1000 (1x10-3)-2 = divide the Factor by 100 (1x10-2)-1 = divide the Factor by 10 (1x10-1)0 = multiply the Factor by 1 (1x100)1 = multiply the Factor by 10 (1x101)2 = multiply the Factor by 100 (1x102)Least Significant Digit(LSD)Set the graduation.Settings of 0.5, 1, 2, 5, 10, 100 are available.The Custom Unit’s name can be customized up to 3 characters.Note: Custom Unit is locked at Off position when the Security Switch is set to the locked position.Custom Unit is not available when Range is set to Dual interval.Set the status.Off = DisabledOn = Enabled2.8 GMP MenuEnter this menu to set the Good Manufacturing Practice (GMP) data.2.8.1 Date FormatSet the date format.MM/DD/YYYY = Month.Day.YearDD/MM/YYYY = Day.Month.YearYYYY/MM/DD = Year.Month.Day2.8.2 DateSet the date.00 to 9999 = year position01 to 12 = month position01 to 31 = day positionRefer to Section 3.2 Menu Navigation to enter settings.2.8.3 Time FormatSet the time format.24 hr = 24 hour format.12 hr = 12 hour format.