TC--C2 COMPACT TROLLEY74.7 Four foot holders should be affixed to the second and third shelves in suitable positions for the equipment being used.4.8 Place items of equipment onto the shelving, ensuring that the equipment feet are located in the foot holders to preventequipment from moving.D Refer to individual equipment instructions for connection details. Cable restraints are provided on the rear of eachshelf.D The TC--C2 has been designed to allow use of the Olympus MB--631 isolation transformer, which should be seatedin the base of the TC--C2 towards the rear.4.9 Plugblock versions only: Using a suitable flat bladed implement, turn the fastener 90˚ and release the plugblock cover.Connect equipment power cables to the plugblock ensuring they exit from the base allowing the cover to be closed. Tidyexcess lengths of cable using the cable straps on the rear of the workstation. Replace the cover and lock the fastener.Cable strapsPlugblock coverFastenerFIGURE 4-5Connection informationWhen assembling equipment onto the workstation, the installation should meet the requirements of EN IEC 60601--1--1,the safety standard for medical electrical systems.This standard requires the aggregate earth leakage current of items powered via a plugblock to be less than 0.5mA.This may be achieved by limited use of the plugblock or the use of an isolation transformer.Before commissioning an installation where an isolation transformer is included, ensure its rating will not be exceededby the equipment connected to it.The plugblock is provided with a cover to prevent connection of equipment that may cause the earth leakage current toexceed IEC limits.The plugblock must not be placed on the floor when in use.The maximum permitted load on the plugblock is 15 amps in USA and 13 amps in Europe.The plugblock must only be used to power equipment which forms part of the System.Equipment powered through an Isolation Transformer must not be connected to equipment not meeting the leakagecurrent requirements of EN IEC 60601--1 which is connected directly to the wall socket as this may compromise systemsafety.The connection of equipment to the plugblock subsequent to the initial system installation must be undertaken bysuitably qualified personnel in order to minimise the risk of system safety being compromised by excessive leakagecurrents.VISY218