CMC 356 Reference Manual1229.6.10 C-ShuntC-Shunt 1 and C-Shunt 10 are precision shunts for current measurements.They can be directly inserted into the binary/analog inputs of a CMC 356with ELT-1 option. The ELT-1 option enables the CMC 356 to measureanalog signals; see 6.10 ”Option ELT-1” on page 67.Figure 9-21:C-Shunt 1Order numbers:C-Shunt 1: VEHZ0080C-Shunt 10: VEHZ0081C-Shunt 10 is a 10 mΩ precision shunt for current measurements. It can bedirectly inserted into the test set’s binary/analog inputs.C-Shunt 1Electrical resistance 0.001 ΩResistance Tolerance 0.1 %Temperature coeffizient ≥ 30 ppm/K in the range 0 ... +70 °C(32 ... +158 °F); according to IEC 60115-1 4.8Maximum current 32 A continuousInsulation class 600 V CAT IIC-Shunt 10Electrical resistance 0.01 ΩResistance Tolerance 0.1 %Temperature coeffizient ≥ 18 ppm/K in the range 0 ... +70 °C(32 ... +158 °F); according to IEC 60115-1 4.8Maximum current 12.5 A continuousInsulation class 600 V CAT II