CMC 356 Reference Manual305.1.1 Generator Combination Socket for VOLTAGE OUTPUT andCURRENT OUTPUTThe combination socket CURRENT OUTPUT / VOLTAGE OUTPUTsimplifies the connection of test objects to the CMC 356. The three voltageoutputs (VOLTAGE OUTPUT 1-3) as well as the CURRENT OUTPUT A arewired to the combination socket (refer to table 5-1 on page 31).Figure 5-4:Generator combinationsocketWARNING:The connections on the socket are dangerous when thetest set is turned on.Follow the safety information provided at the beginning ofthis manual when connecting the generator combinationsockets.If a dangerous voltage (greater than 42 V) is applied to the socket, awarning indicator lights above the socket.For currents greater than 25 A, the test object (load) should be exclusivelyconnected to the 4 mm/0.16 " banana sockets and not on the generatorconnection socket.Front view View onto the connector fromthe cable wiring side