In addition to these operations, the analyzer can he preset to any of16 setups. The following describes how to preset the analyzer.NON.OCOCOoPreset40) Preset is a facility that enables to store various setups formeasurements and to recall them with simple keystrokes. Thefollowing are the basic procedure.Basic procedure for storinga. To set up a required state, do normal operations forindividual functions.b. Press the STO (WRITE) key of the MODE select keys(22) ; this will cause- the blinking of two keys £a”Jand (Tj.c. Specify a preset address with a pair of keys(example: [e] [Tj) - Keys [~a] and [T] go off and thespecified keys begin to blink.d. Press the STO (WRITE) key. The specified two keysstop blinking and remain lit. This implies thesetup has been stored.Basic procedure for recallingPress the RCL key of the MODE select keys (22) ; twokeys. Fa” and [T] will be lit. The contents storedat address HQ are recalled.Press the keys for a specific address (example:pjTj) . The selected two keys light, then thecontents at that address are recalled.41) Data on the following items can be stored:a. Frequency range ... Both the oscillator frequency range and thefundamental frequency range for thedistortion measurementb. Frequency .Both the oscillator.frequency and. thefundamental frequency for the distortionmeasurementc. Output amplitude of the oscillatord. Display of the oscillator (frequency or amplitude)4-19