4 G- ft 8-0 2 0 2OSCOUTPUTVP-7721AINPUT SwitchINPUT—O OUTPU^"Device under test(Non-floating output)—--—--O—Fig. 4-11 Non-floating connectionThe maximum floating voltage is 150 volts. It is specified as thepeak value of the voltage that appears between center conductor ofthe INPUT A or B connector and the common ( ). Accordingly itis equivalent to the maximum allowable input voltage described inthe succeeding paragraph (5). Note that excessive floating voltagemay cause damage to the instrument.- -WARNING -“—-—-If the input signal level or the floating voltage exceeds 50V,particular attention should be paid not to touch an exposed point onthe line to be connected to the center conductor of the INPUT A or Bconnector.Non-floating connection can only be made to the INPUT Aconnector (Fig. 4—11).por floating connection, both A and B connectors are used(Fig. 4-10).The connectors A and B are interchangeable.Note 1)y-b4-25