8-0 2 0 2Table 4-8 S/N measurement limit for floated input(V in rms, values with dB are S/N)Input >3V IV 0.3V Detectionbandwidth BWS component Es>lV lV>Es>0.3V 0.3V>Es>0. IV5HZ-15.99kHa95-dB ^ 85dB 80dB n 80kHz BW85dB 80dB 70dB 500kHz BW16kHz-159.9kHz 80dB 75dB 60dB 500kHz BWThe detection bandwidth is set to 500 kHz by turning off the HPF andboth LPFs, and the detection bandwidth LPF is set to 80 kHz byturning on only the 80 kHz BPF.4.9 Preset ApplicationThe application of the preset operation described in steps 40) to 48) inSection 4.3 to the measurement in Sections 4.5 to 4.8 will result inhighly efficient measurement using the preset function. Up to 16setups, which can be changed easily, can be recalled instantly. Thefollowing remote control expands the number of sets from 16 to 32.4.10 Remote Controla) The following operations can be controlled remotely.(1) Preset recall operation (when the LED in the RCL key of the MODEselect key (^22) is on)Recall operation of 16 preset setups by eight keys (equiva¬lent to keys A to D, 1 to 4 on the instrument).. Recall operation of additional 16 setups by four keys above(equivalent to keys A to D) and four additional keys.(2) Selection of digital display (frequency or output amplitude).Equivalent to the DISPLAY key (?) on the instrument.(3) Modify operation (frequency or output amplitude)Equivalent to the operation of the' MODIFY control @ • Thedisplay selected by item 12) above can be incremented. Thedigit specified by the MODIFY digit select switch (27) iscontrolled.(4) Filter selectionEquivalent to the HPF 400 Hz (S) , LPF 30 kHz (20) , 80 kHz(19) , and WEIGHTING @ keys.4-34