QIAsymphony SP/AS Consolidated Operating Guide 12/2017 183ImportantWhen using the QIAsymphony Cabinet SP/AS, the waste bin should be emptiedto avoid contamination inside the cabinet.Important Check the waste bin regularly.Important Residual liquid from the tip disposal chute may drip inside the cabinet.For detailed information, refer to the QIAsymphony Cabinet SP/AS User Guide.14.3.2 Regular maintenance procedure for the QIAsymphony SP1. Remove and safely store eluates from the “Eluate” drawer. As an optional step, download theresult file(s) and ensure that the files have been backed up.2. Remove used sample tubes/plates from the “Sample” drawer and discard according to yourlocal safety regulations.3. Remove reagent cartridges from the “Reagents and Consumables” drawer. Seal partially usedreagent cartridges and store according to the instructions in the handbook of theQIAsymphony Kit you are using. Discard used reagent cartridges according to your localsafety and environmental regulations.4. Replace the tip disposal or waste bin bag if it is full.5. Close unit boxes filled with waste plasticware and discard according to your local safetyregulations.6. Check the magnetic-head guards.ImportantWhen using the QIAsymphony Cabinet SP/AS, the waste bin should be emptiedto avoid contamination inside the cabinet.For detailed information, refer to the QIAsymphony Cabinet SP/AS User Manual.If required, clean the magnetic-head guards before starting the next protocol run. Proceed asfollows:1. Open the Maintenance SP menu and run the service protocol Magnetic head guards. Gentlyraise the catches to release the magnetic-head guards.2. Wipe the magnetic-head guards with ethanol-based disinfectant (e.g., Mikrozid), andincubate as appropriate.3. Wipe with a lint-free cloth moistened with water and wipe dry with paper towels. Replace themagnetic-head guards.