5.DISPLAY SETTINGS565-9. One-Touch Tare SubtractionThis function clears Net weight by equalizing Gross weight and Tare.Note this function is activated only if reading is stable: stability criteria aredefined in Restriction on Tare Subtraction Function.Allowed range of Tare subtraction: All range or zero < Tare <= Capacity.* In case where Net weight is not zero-cleared after One-Touch Tare Subtraction,check the following:※ Tare subtraction is reset when turning off power.5-10. One-Touch Tare Subtraction ResetTare subtraction can be reset by the user. Resetting tare subtraction zero-clearsthe set value for Tare, equalizing Net weight and Gross weight.* If Net weight and Gross weight are not equalized by resetting tare subtraction,check the following.Cause Suggested ActionGross weight is displayedNet weight is displayed in n+4, n+5 CH.Gross weight is displayed in n+2, n+3 CHMake sure you are checking correct area.Reading is not stable(check stability criteria)Make sure reading is stable: “1” on the 10th bit of n+5CH indicates that reading satisfies stability criteria.Check this bit before trying to clear Net weight.Reading is outside the allowedtare subtraction range (checkrange setting)Use this function while reading is within the allowablerange for tare subtraction.Cause Suggested ActionDigital tare subtraction isenabledZero-clear the set value for Tare, or disable Digitaltare subtraction.