143 Product descriptionVEGABAR 82 • Slave sensor for electronic differential pressure48046-EN-190320The front-flush, hygienic installation of the measuring cell acc. to 3Ais particularly suitable for food applications. The sealings are installedgap-free. The additional front sealing for the measuring cell protectsalso the glass joint. A hole in the process fitting is used for leakagedetection.16 2345Fig. 9: Hygienic installation of the measuring cell acc. to 3-A (example: Clampconnection)1 Measuring cell2 Seal for the measuring cell3 Process fitting4 Additional, front seal for measuring cell5 Diaphragm5 Hole for leakage detection3.3 Supplementary cleaning proceduresThe VEGABAR 82 is also available in the version "Oil, grease andsilicone-free". These instruments have passed through a specialcleaning procedure to remove oil, grease and paint-wetting impair-ment substances (PWIS).The cleaning is carried out on all wetted parts as well as on surfacesaccessible from outside. To keep the purity level, the instruments areimmediately packed in plastic foil after the cleaning process. Thepurity level remains as long as the instrument is kept in the closedoriginal packaging.Caution:The VEGABAR 82 in this version may not be used in oxygen ap-plications. For this purpose, instruments are available in the specialversion "Oil and grease-free for oxygen applications".3.4 Packaging, transport and storageYour instrument was protected by packaging during transport. Itscapacity to handle normal loads during transport is assured by a testbased on ISO 4180.The packaging of standard instruments consists of environment-friendly, recyclable cardboard. For special versions, PE foam or PEfoil is also used. Dispose of the packaging material via specialisedrecycling companies.Transport must be carried out in due consideration of the notes on thetransport packaging. Nonobservance of these instructions can causedamage to the device.The delivery must be checked for completeness and possible transitdamage immediately at receipt. Ascertained transit damage or con-cealed defects must be appropriately dealt with.Installation in hygienicfitting acc. to 3-APackagingTransportTransport inspection