If there is no electronics module available on site, one can be orderedfrom the VEGA agency serving you.The new electronics module must be loaded with the order data of thesensor. These are the options:l At the factory by VEGAl Or on site by the userIn both cases, the sensor serial number is needed. The serial numbersare stated on the type label of the instrument or on the delivery note.Information:When loading on site, first of all the order data must be downloadedfrom the Internet (see operating instructions manual "Oscillator").The electronics modules are adapted to the respective sensor anddiffer in their signal output or in their power supply. You can find asuitable electronics module in the following overview.The oscillators differ only in their signal output and are suitable for allseries 60 sensors.The following types are available:l CL-E60H (4 … 20 mA/HART)l CL-E60P (Profibus PA)l CL-E60F (Foundation Fieldbus)In Ex applications only one electronics module with respective Exapproval may be used.8.4 Shortening the electrodeThe probe (rod) can be shortened by any length.1 Shorten the rod with a cut-off wheel or metal saw at the lower end.Make sure the length is correct before shortening.2 Carry out a fresh adjustment8.5 Instrument repairIf a repair is necessary, please proceed as follows:You can download a return form (23 KB) from our Internet homepagewww.vega.com under: "Downloads - Forms and certificates - Repairform".By doing this you help us carry out the repair quickly and withouthaving to call back for needed information.l Print and fill out one form per instrumentl Clean the instrument and pack it damage-proofSensor serial numberAssignment40 VEGACAL 62 • Profibus PA8 Maintenance and fault rectification30025-EN-090603