R Series Service Manual9650-0903-01 Rev. L A-3Delivered Energy at Every DefibrillatorSetting into a Range of Loads* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is greater.SelectedEnergyLoad25 50 75 100 125 150 175 Accuracy*1 J 1 J 1 J 1 J 1 J 1 J 1 J 1 J ±15%2 J 1 J 2 J 2 J 2 J 2 J 2 J 2 J ±15%3 J 2 J 3 J 3 J 3 J 3 J 3 J 3 J ±15%4 J 3J 4 J 4 J 5 J 5 J 5 J 4 J ±15%5 J 3 J 5 J 6 J 6 J 6 J 6 J 6 J ±15%6 J 4 J 6 J 7 J 7 J 7 J 7 J 7 J ±15%7 J 5 J 7 J 8 J 8 J 8 J 8 J 8 J ±15%8 J 5 J 8 J 9 J 9 J 10 J 9 J 9 J ±15%9 J 6 J 9 J 10 J 11 J 11 J 11 J 10 J ±15%10 J 7 J 10 J 12 J 12 J 12 J 12 J 12 J ±15%15 J 10 J 16 J 17 J 18 J 18 J 18 J 17 J ±15%20 J 14 J 21 J 23 J 24 J 24 J 24 J 23 J ±15%30 J 21 J 32 J 35 J 36 J 37 J 36 J 35 J ±15%50 J 35 J 54 J 59 J 61 J 62 J 61 J 59 J ±15%70 J 49 J 76 J 83 J 85 J 87 J 86 J 83 J ±15%75 J 53 J 81 J 89 J 91 J 93 J 92 J 89 J ±15%85 J 60 J 92 J 101 J 104 J 106 J 104 J 101 J ±15%100 J 71 J 109 J 119 J 122 J 125 J 123 J 119 J ±15%120 J 85 J 131 J 143 J 147 J 150 J 147 J 143 J ±15%150 J 107 J 164 J 180 J 183 J 188 J 184 J 179 J ±15%200 J 142 J 230 J 249 J 253 J 269 J 261 J 260 J ±15%