R Series Service Manual1-22 9650-0903-01 Rev. L11.0 Defibrillator Self TestSHOCK HAZARD!TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU STARTCONDUCTING THE DEFIBRILLATOR TESTS.Keep hands and all other objects clear of the multi-function cable connections and defibrillator analyzer when dischargingthe defibrillator.Before you discharge the defibrillator, warn everyone near the equipment to STAND CLEAR.Caution Do NOT internally discharge the unit more than 3 times in 1 minute. Note that multiple rapidly repeating internal discharges at more than30 Joules may damage the unit.Tools Needed:• ZOLL Medical Electrode Connector (Fluke Biomedical Part Number 3010-0378 or equivalent).• Impulse 4000 or equivalent defibrillator analyzer.• ECG Cable.• Stop watch.Test Setup:1. Ensure the unit is turned off.2. Ensure the ECG cable is connected to the R Series unit and the defibrillator analyzer.Note: The OneStep cable should not be connected to any equipment at the beginning of this test.Do this... Observe this... Pass/Fail11.1 Turn the selector switch to DEFIB mode (ForAED units, turn to ON, and select MANUALMODE.)Set leads to PADS.CHECK PADS/POOR PAD CONTACT message displays. o o11.2 Connect the OneStep cable to the test port onthe right side of the R Series.DEFIB PAD SHORT message displays. o o