R Series Service Manual1-28 9650-0903-01 Rev. L15.0 Summary Report TestTools Needed: Impulse 4000.Test Setup:1. Connect the OneStep cable to the Impulse 40002. If you are using paddles, place the paddles on the analyzer’s discharge plates.Do this... Observe this... Pass/Fail15.1 Press the Report Data * softkey, press Erase,then Erase All.ERASING REPORT displays. Wait for the message to clear. o o15.2 Set selector switch to DEFIB (for AED units turnto ON, and select MANAUL MODE.) Select200J using the ENERGY SELECT button, andpress the CHARGE button. When charged,press the SHOCK button to discharge into thedefibrillator analyzer.The unit successfully discharges and if configured to do so, prints a strip chart. o o15.3 Wait 18 seconds, then press the Code Markersoftkey. Press the CPR softkey. Wait 20seconds and turn unit off. (This will allow codemarker to be saved in the summary.)The Code Markers display. (After CPR Code Marker is selected, the unitautomatically exits the Code Marker options menu.) o o15.4 Verify that the unit has been off for a minimumof 10 seconds and then turn the unit on. Pressthe Report Data * softkey, then press PrintChart, then Print All.Summary report prints. The report displays the correct date, time, the shockdelivered, and Code Marker event. o o* For software versions 3.0 or earlier, the unit displays Report instead of Report Data.Record your results on the Maintenance Test Checklist.