34 New installation of S8 / S 81 ceiling mountM-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0Printed on 19. 11. 2007Installing the conduitsThe wiring can be run either above the false ceiling or in conduits installedin the structural ceiling. There are two options for the installation of con-duits in the structural ceiling.e The exit of the conduits is positioned beside ceiling anchor plate (8).e The exit of the conduits is positioned at the center of ceiling anchorplate (8).The conduits have an internal diameter of 22 to 50 mm (0.9 - 2").A conduit with a minimum internal diameter of 35 mm (1.4") is required forthe MediLive video line.Key1 Power line3 x 1.5 mm 22 Wall socket for foot control panelAs required.3 Potential equalizationAs required.4 TV internal diameter 50 mm (2")As required.5 OR illuminators24 V DC with backup power supply, customer's responsibility (as re-quired)6 Other OPMI accessoriese. g. BIOM/SDI, additional illumination, AUX output port, etc. (as re-quired)7 Further conduitsWe recommend installing further conduits for future upgrades.