IntroductionSetupSoftwareMaintenanceSpare Parts Specifications41MaintenancePerformingSimulations5. Remove urine reservoir.6. Lay new reservoir in position in reservoir bay, with red filler capon the manikin’s left, facing upwards.7. Connect yellow Luer on urine reservoir inlet tube to yellow urinepressurization outlet on bulkhead. (3/4 turn, do not overtighten).8. Connect white Luer on urine reservoir outlet tube to white Lueron urine connector tube.9. Ensure slide clamp is open.10. Replace reservoir cover and reconnect retaining loop and Velcro®strap.Blood Reservoir ReplacementThe blood reservoir is located in the reservoir bay, beneath the urinereservoir.To replace blood reservoir:1. Remove reservoir cover and urine reservoir.2. Clamp blood outlet with clamp.3. Undo blue Luer connector from the blue bulkhead fitting.4. Remove blood reservoir.5. Lay new reservoir in position in reservoir bay, with red filler cap tothe manikin’s right, facing upwards.6. Connect blue Luer connector to blue (blood pump inlet)bulkhead fitting.7. Ensure slide clamp is open.8. Replace urine reservoir.9. Replace reservoir lid.Pelvis ReplacementTo detach and replace the pelvis:1. Turn the manikin over so that the backside is facing up.2. Using a Phillips head screw driver, unscrew the access panel andremove.3. Locate the black electrical connector and two clear tubes, anddisconnect them.– Disconnect the clear tubes by twisting the fittings.– Disconnect the electrical connector by depressing the blackbutton on the side of the connector.4. Turn the manikin back over.5. Remove abdominal skin.6. Using a 7/16 wrench and a Phillips head screw driver, remove thebolts on both sides of manikin.