page 26 of 36Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefullythe validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011ParameterXX:XXXXXXX:XXXXXXX-1FunctiontimealternativelyCommandOUT_SP_1OUT_SP_2OUT_VENTOUT_SENSORREMOTEOUT_SP_11121314OUT_SP_21222324Write commandsDescriptiontime interval for automatic sending of allpressure values mm:ssssstime interval for recording*0: venting valve closed1: venting valve open2: venting until atmospheric pressure1-8 external sensors (if connected)switch off deviceswitch on deviceThe single abbreviations of a command are separated by underscores (ASCII 5FH).The string is terminated with (ASCII 0DH, ASCII 0AH).Connector assignment2: RxD 5: Mass3: TxD 7: RTS4: DTR 8: CTS9: +5V (Bluetooth)* When changing the time interval, recording starts anew, existing data are deleted. Time interval ”0”stops recording, data are stored.setpoint On 1setpoint On 2setpoint On 3setpoint On 4setpoint Off 1setpoint Off 2setpoint Off 3setpoint Off 4XXXX.X mbar/Torr/hPaX.XXE±X mbar/Torr/hPaXXXX.X mbar/Torr/hPaX.XXE±X mbar/Torr/hPa