13System User Guide 4500 Series of InstrumentsRUO-IDV-05-0264-A 119 of 138Service and Maintenance InformationRegularly clean and maintain the system for optimal performance. For information on tuningfrequency, refer to Table B-1. For information on maintaining the ion source, refer to the ionsource Operator Guide.Table 13-1 provides a recommended schedule for cleaning and maintaining the system. Contacta Qualified Maintenance Person (QMP) to order consumable parts. Contact an AB SCIEXrepresentative for maintenance service and support.WARNING! Potential Radiation Hazard, Biohazard, or Toxic ChemicalHazard: Determine whether instrument decontamination is required priorto cleaning. Decontamination should be conducted prior to cleaning ifradioactive materials, biological agents, or toxic chemicals have beenused with an instrument.Table 13-1 Maintenance TasksComponent Frequency Task For more information...Curtain plate Daily Clean Refer to Clean the Curtain Plate onpage 124.Orifice plate(front)Daily Clean Refer to Clean the Front of the OrificePlate on page 124.Orifice plate(front and rear)As needed Clean Contact the local QMP or AB SCIEX FieldService Employee (FSE).QJet ® ion guide As needed Clean Contact the local QMP or FSE.Q0 and IQ1 lens As needed Clean Contact the local QMP or FSE.InstrumentsurfacesAs needed Clean Refer to Clean the Surfaces on page 120Drain bottle As needed Empty Refer to Empty the Source Exhaust DrainBottle on page 125Roughing pumpoilAs needed Check and fill Contact the local QMP or FSE.Instrument airfilterEvery 6monthsInspect andclean or replaceContact the local QMP or FSE.Roughing pumpsoilLeyboldSogevacSV28: Every 6monthsVarian MS 40:AnnuallyReplace Contact the local QMP or FSE.Electrode As needed Inspect andclean or replaceTurbo V ™ Ion Source Operator Guide.Coronadischarge needleAs needed Replace Turbo V™ Ion Source Operator Guide.