32 TZIDC DIGITAL POSITIONER | OI/TZIDC-EN REV. B… 7 CommissioningSample parameters‘Change the zero position of the LCD display from clockwise(CLOCKW) to counter-clockwise limit stop (CTCLOCKW)’Initial situation: the positioner is in bus operation on theoperating level.1. Switching to the configuration level:• Press and hold down© and ª simultaneously,• additionally quickly press and release ENTER,• Wait for the countdown to go from 3 to 0,• Release© and ª.The following is now shown in the display:M10906i2. Switching to parameter group 3._:• Press and hold downMODE and ENTER simultaneously,• additionally quickly press and release © 2×,The following is now shown in the display:M10906j• ReleaseMODE and ENTER .The following is now shown in the display:M10906k3. Selecting parameter 3.2:• Press and hold downMODE,• additionally quickly press and release © 2×,The following is now shown in the display:M10906l— Release MODE.4. Changing parameter settings:• Quickly press and release © to select CTCLOCKW.5. Switching to parameter 3.3 (Return to operating level) andsaving the new settings:• Press and hold downMODE,• additionally quickly press and release © 2×,The following is now shown in the display:M10906m• ReleaseMODE,• Quickly press and release © to select NV_SAVE,• PressENTER and hold down until the countdown goesfrom 3 to 0.The new parameter setting is saved and the positionerautomatically returns to the operating level. It continues in theoperating mode that was active prior to the configuration levelbeing called up.Setting the option modulesSetting the mechanical position indication1. Loosen the screws for the housing cover and remove it.2. Rotate the position indicator on the shaft to the desiredposition.3. Attach the housing cover and screw it onto the housing.Tighten the screws so that they are hand-tight.4. Attach the symbol label to mark the minimum and maximumvalve positions on the housing cover.NoteThe labels are located on the inside of the housing cover.