19-2 Art: 714381-01J Rev. Date: 11-Mar-14STARTUP MESSAGESOverview Whenever the analyzer is turned on using the On/Off key, the analyzer performsself-checks. If a condition that should be corrected in the near future, but thatwill not affect results is detected, a warning is displayed. The operator pressesthe 1 key to continue with testing. If the analyzer has been customized to disabletesting under any of these conditions, the condition must be corrected andthe analyzer turned off and back on before testing will be enabled.Message on Display Explanation How to RespondElectronic Simulator TestRequiredAnalyzer customized to alertthe operator that a scheduledsimulator test is due.Insert the external ElectronicSimulator at the earliestconvenient time.Stored Memory Low Memory space for 50 unsenttest records available before the“Stored Memory Full” messageis displayed.Place the analyzer in aDownloader.Stored Memory Full The analyzer is customizedto alert the operator that thememory for unsent recordsis full. If the operator doesnot transmit the test recordsto the Point-of-Care CentralWorkstation, the analyzer willeither block further testing orwill overwrite oldest recordsdepending on how the analyzeris customized.Place the analyzer in aDownloader.Upload Required The analyzer is customizedto alert the operator that ascheduled transmission of testrecords to the Central DataStation is due.Place the analyzer in aDownloader.Battery Low Battery voltage has droppedto 7.4 volts. There is sufficientpower to test a few morecartridges, the numberdepending mainly on the typesof cartridges in use. Under thiscondition, a flashing batteryicon will also appear on theresult page, the Test Menuscreen, and the AdministrationMenu screen.Change the disposable lithiumbatteries or recharge therechargeable battery.CLEW Expiring, UpdateRequiredMessage appears 15 daysbefore the software expires.Update the analyzer before theexpiration date.