Azure™ cSeries User ManualPage 6Table of SpecificationsImage Resolution 5.4M pixels (c150, c200); 8.3M pixels (c300, c400, c500, c600)Epi-illuminations White LED; Blue LED; RGB LED (c400, c600); 660nm LD (c500, c600)785nm LD (c500, c600)Trans-Illumination 302nm/365nmLens 25mm/F0.95 motorized (c300, c400, c500, c600); 8mm/F1.2 (c200)Maximum Field of View 20 cm x 15 cmImage output 16 bit TIFF, JPEGWorking environment Ambient temperature/humidity: 0 –30°C / 85%Power requirement 100-240 VAC, 4ADimensions (W x D x H) 38 cm x 30 cm x 45 cmWeight 22 kg (49 lbs.)2. Installation and Setup2.1 System PlacementWarning! Excessive Weight Hazard – Please use two or more people to lift the system. Failure todo so can result in system damage and personal injury.As with all electrical instruments, the Azure cSeries imaging system should be located away from water,solvents, or corrosive materials, on a flat and stable surface with adequate clearance on all sides. Thesystem must remain stationary during operation.Further, the system should be placed away from interfering electrical signals and magnetic fields. Ifpossible, a dedicated electrical outlet should be used to eliminate electrical interference from otherinstrumentation in your laboratory.The Azure cSeries imaging system should be installed at no more than 3000 meters above sea level.2.2 Connecting to PowerThe power entry module is located in the lower right on the back panel of thesystem. Connect the power cord to a secure power outlet.It is important to connect the system to a well grounded power source. AzureBiosystems recommends that you employ a surge protecting power strip toprotect against potential electrical component damage from power surges.2.3 Power On/Off the SystemPower on the system by:1. Using the power switch on the back of theinstrument.2. Pressing the power button on the top of the tablet.