Page 28 of 110Connection to Device Under Test (DUT)Figure COI-1 illustrates the 1910 Inductance Analyzer connected to a device under test using the4-BNC to 2-Kelvin Clip lead set (QT P/N 1700-03).10IET1910 Inductance AnalyzerSTOP STARTFAIL PASSILUTILITYPL PH IHPROGRAMBias OnRemoteLockoutIET LabsModel 1910 V1.4Chassis1700-03HCURH POTLPOTL CURH CURHPOTLCURL POT(+)(-)DUTFigure COI-1: Connection to DUT using 1700-03 Kelvin LeadsConnection to the Device under Test (IET Labs. Cables)Connection as labeledon 1920 front panelConnections withcables labeledConnections withcables color codedConnections withcables color codedPH connects to PH or HS red/white redIH connects to IH or HD red redPL connects to PL or LS black/white blackIL connects to IL or LD black blackInitiating Measurements1) The instrument must be in the READY mode, or displaying the results from a previousmeasurement before initiating a test1 Ls Rs 1.0000kHz1.000V No Bias Auto High2) Test conditions and Utility functions should be programmed as previously described inthis section.ORSelect the test number by pressing the Up or Down arrow key, if known test conditionsare already stored in internal memory.6 Prim Param ReadyAutomatic