Operation/Programming Page 45 of 110Up or Down arrow keyto change voltageRight arrow key to programnext parameterorPROGRAM key to exit ProgramMode and return to Ready mode.2.3.5 Bias CurrentThis function allows the user to program DC bias current to Off or from 1mA to 1A in 1mAsteps with a maximum compliance voltage of 10V, i.e. (I) times (R) must be less than 10 volts.When selected for Off no bias current is applied to the device under test, when programmed for abias current this DC current flows to the device under test.NOTE:When using the 1910 (in external bias mode) with the IET Labs. 1320 DC Bias Current Source, the Bias Currentmust be set to OFF and the 1320 set to Handler mode. Refer to the 1320 instruction manual. Connect the 1910input BNC’s (4) to the 1320 FROM L METER BNC’s (4), using BNC to BNC cable. Connect 1910 rear panelREMOTE I/O to the 1320 HANDLER INTERFACE using a 1320-07 interconnect cable.Up or Down arrow keyto change bias current1 Amplitude Program1.0000 V2 Amplitude Program1.00 mV1 Amplitude Program1.0000 V1 Bias Current ProgramOff1 Bias Current Program1.00 mA1 Bias Current Program1.0000 A1 mV to 1 V1 mA to 1 A