30 AEJ_N/AES_N-BA-e-13218 MenuThe menu is subdivided into seven menu blocks (P1 – P7).P1 CAL [adjustment]P1.1 iCAL | [internal adjustment] ALJ onlyP1.2 ECAL | [external adjustment] *P1.3 tCAL | [adjustment test]P1.4 ACAL | both/nonE/tenno/tinnE [automatic adjustment] ALJ only *P1.5 CALt | 1 h ÷ 12h [time setting for automatic adjustment] ALJ only *P1.6 CALr | YES/no [printout adjustment log]* = locked in devices with type approval certificateP2 GLP [good lab practice]P2.1 USr | _ [Enter user name]P2.2 PrJ | _ Enter [project name]P2.3 Ptin | YES/no [printout time]P2.4 PdAt | YES/no [printout date]P2.5 PUSr | YES/no [printout user]P2.6 PPrJ | YES/no [printout project]P2.7 PId | YES/no [printout serial number weighing balance]P2.8 PFr | YES/no [frame printout]P3 rEAd [main functions]P3.1 AuE | Stand/Slouu/FASt [filter settings]P3.2 ConF | FASt_rEL/Fast/rEL [rest control display]P3.3 Auto | On/OFF [Auto zero]P3.4 Ldi9 | ALuuAYS/never/uu_StAb [delete last decimal place]P4 Print [parameter for serial interface RS 232]P4.1 bAud | 2400/4800/9600/19200 [Baud rate]P4.2 CntA | YES/no [continuous output standard weighing unit]P4.3 Cntb | YES/no [continuous output currently set weighing unit]P4.4 rEPL | YES/no [manual (press key) or automatic output]P4.5 PStb | YES/no [output stable/instable weighing values]P4.6 Lo | 000.0000 [input minimal weight for automatic output]