AEJ_N/AES_N-BA-e-1321 8316 Instant helpPossible causes of errors:In case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.Fault Possible causeThe displayed weight doesnot glow.• The balance is not switched on.• The mains supply connection has been interrupted(mains cable not plugged in/faulty).• Power supply interrupted.The displayed weight ispermanently changing• Draught/air movement• Glass doors not closed• Table/floor vibrations• Weighing pan has contact with other objects.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)The weighing result isobviously incorrect• The display of the balance is not at zero• Adjustment is no longer correct.• The balance is on an uneven surface.• Great fluctuations in temperature.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)Automatic adjustmentcarried out frequently.• Severe temperature variations in the room or theinstrumentNo data transfer betweenprinter and balance. • Communication settings are wrong.The menu setting cannot bechanged.• Menu item is locked for models with type approvalcertificate.