36 AEJ_N/AES_N-BA-e-1321 Confirm your selection by pressing the PRINT-key; weighing scale will return tomenu. For returning to weighing mode with set weighing unit, see chap. 8.2.9.4 Weighing units switch-overA weighing unit may be changed by pressing the UNIT-key several times:[g] → [mg] → [ct] → [oz] → [ozt] → [dwt] → [t] → [mom]→ [G] → [g]Note:In calibrated model selection is restricted to options [g] → [mg] → [ct].9.5 TaringThe dead weight of any weighing container may be tared away by pressing a button, so thatthe following weighing procedures show the net weight of the goods to be weighed. Put vessel of goods to be weighed on weighing plate and close the windscreen doors. Wait until the stability display appears ( ), then press TARE. The weight ofthe container is now internally saved. Weigh the goods to be weighed and close the wind screen doors. Wait until the stability display appears ( ). Read net weight.• When the balance is unloaded the saved taring value is displayed withnegative sign.• To delete the stored tare value, remove load from weighing pan andpress TARE.• The taring process can be repeated any number of times. The limit isreached when the whole weighing range is exhausted.