AEJ_N/AES_N-BA-e-1321 7314.3 Menu settings “P4 PRINT“Menu call “P4 PRINT“, see chap. 8.1.Menu overviewFunction Settings Description of optionsP4.1 bAud 2400 bit/sTransfer speed4800 bit/s *9600 bit/s19200 bit/sP4.2 CntAno* Continuous output disabled in standardweighing unit,Output requires key operationYES Continuous output in standard weighing unitP4.3 Cntbno* Continuous output disabled in current weighingunit, output requires key operationYES Continuous output in current weighing unitP4.4 rEPLno* Manual issue after pressing the PRINT key.YES Automatic issue of first stable weighing valueSequence of operations:1. Taring2. Place weight, issue of first stable weighingvalue3. Renewed output only possible after weightwas removed. Prerequisite: Display +/- 50display steps from zero.4. Place next weight.P4.5 PStbLocked in models withtype approval certificateno Output even for unstable weighing valueYES* Output for stable weighing value onlyP4.6 Lo 0.01* Input minimum weight for automatic output:Weighing value issued automatically if currentweighing value exceeds entered minimumvalue. The next weighing value will not beissued unless the weighing value hasmeanwhile dropped below the entered weighingvalue.Factory settings are marked by *.