CB/FCB-BA-e-0640 238.2 Leave menu structureEverywhere in the menu it is possible to leave the menu structure and thereby saveor reject the changes made.After pressing the key „Exit“ is displayed.A : Use the (YES) key, to confirm. After that „store“ is displayed. If it shall besaved, press the key repeatedly.If the menu shall be left without saving,press the NO key.B : The (NOT EXIT) key must be pressed, if the next menu item shall bereached. After having set all the individual adjustments, it can be saved.8.3 Dosing and Zero-trackingThe Auto-Zero function is used to tare small variations in weight automatically.In the event that small quantities are removed or added to the material to beweighed, incorrect weighing results can be displayed due to the “stabilitycompensation“ in the balance. (Example: Slowly draining fluids from a container onthe balance).When apportioning involves small variations of weight, it is advisable to switch off thisfunction.If Zero-Tracking however is switched off, the weighing display becomes more busy.Activate/deactivate Zero-Tracking Balance display1. Keep the key pressed until „Unit" is displayed. Unit2. Press the key several times until „tr“ is displayed. tr3. Press the key to activate the function. tr on4. By pressing once more the key, the function isdeactivated. tr off5. The changed setting is taken over by pressingthe key.6. The balance returns to the weighing mode. 0.0 g