CB/FCB-BA-e-0640268.6 Selection printed editionUse this function to select which data are sent via the RS232C interface.To achieve this, follow the sequence of operations below: (is not valid for the datatransfer mode BAPr )Switch-on the balance using the key and wait for the „0“ display.Press the key and keep it pressed until „UNIT“ appears on the display.Press the key twice, in the display appears „LAPr“.Confirm by pressing the key.Now it is possible to scroll forward in the available edition parameters using thekey, to scroll backward, use the key.After actuating the key, the current state is displayed ( on / off ).Use the or the key to change the state.After that the current state is saved using the key and the menu is left.By that way the user can configurate his own data block, which then is sent to aprinter or to a PC.Display Status Function„Hdr“ On / Off Edition of the headlines„GrS“ On / Off Edition of the total weight„Net“ On / Off Edition of the net weight„tAr“ On / Off Edition of the tare weight„N7E“ On / Off Edition of the stored weight„PCS“ On / Off Edition of quantity„AUJ“ On / Off Edition of the unit weight„rqt“ On / Off Edition of the reference quantity„FFd“ On / Off Edition of a page feeding