28 CFS/CCS-BA-e-12207.3 Adjustment bulk material scales CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS30K0.5Not pre-configured by KERNOperation Display Switch-on balance and during the selftest press . Use the number keys to enter password:standard password „0000“ Confirm input by . Use to select bulk material scales („remote“).The appeared [] displays the current balance.Acknowledge with . If necessary, at balance zero display usingselect the weighing unit [kg or lb], which shall be usedfor adjustment. The appeared [] displays the currentweighing unit.Acknowledge with . Ensure that there are no objects on the weighing pan.After successful standstill control press the button. Enter value of the necessary adjustment weight usingnumeric keyboard and acknowledge with .In order to achieve high-quality weighing results in thesense of the measuring technology, it is recommendedto select the nominal value as high as possible.We recommend 80 % max. Place the required adjusting weight carefully in thecentre of the weighing plate and press to confirm. After the adjustment the balance will carry out a self-test.Remove adjusting weight during selftest, balance willreturn into weighing mode automatically.In case of an adjustment error or incorrect adjustingweight the display will show an error message( ), repeat adjustment process.