CFS/CCS-BA-e-1220 31 If necessary, at balance zero display usingselect the weighing unit [kg or lb], which shall beused for linearisation. The appeared [] displaysthe current weighing unit.Acknowledge with . Ensure that there are no objects on the weighingpan. After successful standstill control press thebutton. The weighing value of the first weight will bedisplayed. Place weight 1 carefully in the centre of theweighing plate, wait for the stability display andconfirm by .(example) The weighing value of the second weight will bedisplayed. Remove weight 1 Place weight 2 carefully in the centre of theweighing plate, wait for the stability display andconfirm by .(example) After the adjustment the balance will carry out aself-test. Remove adjusting weight during selftest,balance will return into weighing modeautomatically.In case of fault or a wrong weight an error message( ) will appear, then repeat thelinearization procedure.8.1.1 Table Linearization pointsMax DisplayWeight 1Weight DisplayWeight 2Weight6 kg „A 2 kg“ 2 kg „A 6 kg“ 6 kg15 kg „A 5 kg“ 5 kg „A 15 kg“ 15 kg30 kg „A 10 kg“ 10 kg „A 30 kg“ 30 kg