30 CFS/CCS-BA-e-12208 LinearisationLinearity shows the greatest deviation of a weight display on the scale to the valueof the respective test weight according to plus and minus over the entire weighingrange.If linearity deviation is discovered during a testing instrument control, you canimprove this by means of linearization.Carrying out linearization is restricted to specialist staff possessing wellacquainted with the workings of weighing scales.The test weights to be used must be adapted to the weighing scale’sspecifications; see chapter 3.4 “testing instruments control”.Observe stable environmental conditions. Stabilisation requires a certainwarm-up time.After successful linearisation you will have to carry out calibration;see chapter 3.4 “testing instruments control”.8.1 Linearisation CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 models:Operation Display Switch-on balance and during the selftest press. With the numerical keys input the password „ 9999“:Confirm with . „tECH LoCAL“ is displayed, should it not bedisplayed, with select „tECH LoCAL“.Acknowledge with .The appeared [] displays the current balance.