54 CFS/CCS-BA-e-1220The information if the goods to be weighed are within the tolerance limits, canadditionally be given by an acoustic signal (menu item „ „ see chap. 13.2) For tolerance control, also only one limit value can be set. If both limit values are deleted, the tolerance control is deactivated.12.2.1 Delete limit values:When entering upper and lower threshold, press key and confirm with .12.3 Store dataThe balance has over 100 indirect memories and 5 direct memories (only CFS50K-3 models) for frequently used tare values, average unit weights and additionaltexts. These data can be accessed by calling up the corresponding number of acertain article.12.3.1 Store in indirect memoriesCFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 models:Buttons ~ :Operation Windows If necessary set balance to zero using. Tare, as required.If it’s used as counting system, observe if bulk material or counting balance shallbe tared. Select using bulk material or reference balance accordingly by .The appeared [] displays the current balance.Or place the weighing container and tare with (see chap. 9.4.1) or enter tarevalue numerically (see chapter 9.4.2).Tare values can only be saved when they are situated in the admitted taringrange (factory setting > 2 % max).At values < 2 % max, reset balance with .