PES/PEJ-BA-defsi-05101147.2. Menu overviewThe manufacturer’s setting has a certain standard configuration. This one is markedwith *.Function DisplayorSelectionorDescription of theoptions* 1 Weighing2 Parts counting3 Percent determinationWeighing mode q SET.5 Density determinination of solids* 0 dest. WaterMeasure medium 1q NED. 1 Measuring liquid of your selection* 0 only output measuring value densityData output 12. D.Q.D. 1 Output of all density parameters* 0 OFF (Output only after pressing PRINT key)Density determi-nationAutom.Data output 13. A.Q. 1 ON* 0 Off1 Adding !2C. AD.N."2 Tolerance weighingAdditional functions r SEL3 Combination Toleranceweighing/addingsee chpt. 7.2.10 No zero balancingZero balancing s A.O * 1 Automatic zero balancing is activated.* 23Vibration filter t S.D.4Sensitive and fast (very quiet set-up location).!Robust but slow (very busy set-up location)012Display speed u RE.* 3Setting for dispensingSensitive and fast!Insensitive but slow0 Deactivated* 1 6-digit data format2 7-digit data format see chap. 15.4.1Interface(see chpt. 7.2.1)v K.F.3 extended 7-digit dataformat not documented