PES/PEJ-BA-defsi-051016814.4. Data output14.4.1. Format for data transmissionAny of the data formats below may be set by selecting the relevant function on thebalance ( see menu overview chpt. 7.2):! 6-digit data formatConsisting of 14 words, including final character; CR=0DH, LF=0AH (CR=balancereverse motion / LF=line feed)1 2 3 4 5 6 7 8 9 10 11 12 13 14P1 D1 D2 D3 D4 D5 D6 D7 U1 U2 S1 S2 CR LF! 7-digit data format1 2 3 4 5 6 7 8 9 10 11 12 13 14 15P1 D1 D2 D3 D4 D5 D6 D7 D8 U1 U2 S1 S2 CR LFNOTE: The 7-digit format is identical to the 6-digit, with the exception of the addi-tional character D8.! Extended 7-digit data formatNot documented14.4.2. SignsP 1 = 1 wordP1 Code Meaning+ 2 B H Data is 0 or positive- 2 D H Data is negative14.4.3. Data6-digit data format (D1-D7): 7 words7-digit data format (D1-D8): 8 wordsD1-D7, D8, D9 Code Meaning0 - 9 30 H – 39 H Data 0 to 9 (max. 6 characters in 6-digit format). 2 EH Decimal point, position not fixedSp 20 H Space character, leading cero suppressed/ 2 FH Slash "/" is inserted after the e-value.