PES/PEJ-BA-defsi-05101167.2.1. Parameter additional functionsNot displayed at menu setting „r SEL 0“Function DisplayorSelectionorDescription of options*1 Tolerance marker is always displayed, even ifstandstill control is not yet displayed.Display conditions ofthe tolerance marker 21. CQ.2 Tolerance marker is only displayed in connectionwith standstill control.0 Tolerance marker is only displayed above zerorange (mind + 5).Tolerance range 22. LK.*1 Tolerance marker is displayed for the wholerange.1 1- Limiting point (OK/ -)*2 2- Limiting points (+/OK/-)3 3- Limiting points (1-4)Number of limitingpoints 23. P14 4- Limiting points (1-5)*1 Evaluation for absolute valuesAssessment 24. TXP. 2 Evaluation for difference values (with referenceweight)*0 No signal at limit 1(-)Signal at limit 1 25. BW.K1 Signal at limit 1 (-)*0 No signal at limit 2(Ok)Signal at limit 2 26. BW.21 Signal at limit 2(Ok)*0 No signal at limit 3(+)Signal at limit 3 27. BW.31 Signal at limit 3(+)*0 No signal at limitSignal at limit 4 28. BW.41 Signal at limit 4*0 No signal at limit 5Signal at limit 5 29. BW5 1 Signal at limit 5*1 Display via +, OK or -Display ofResults 2A. LG2For setting 2 limits display in bar graph is possi-ble*1 Permanent output, depending external signalRelay output setting 2B .R.Q.m.2 Output controlled by external signal*1 Adding functionAdd 2C AD.N.2 Adding function with AUTO-TARA