PLE-BA-e-081214⇒⇒⇒⇒⇒ +The weight value of the required adjustment weight will pop up:Place the required calibration weight carefully in the centre of the weighing plate andpress the PRINT key. CAL appears on the display, adjustment is automaticallystarted.Once the adjustment has been completed, unLoAd will appear in the displayTake away adjustment weight• Return to weighing modePress the button repeatedly until the "SAVE"? inquiry appears.Confirm the inquiry by pressing the button to save the changes and to return toWeighing mode.-or-press the button to return to Weighing mode without saving.