PLE-BA-e-0812 2511.3 Auto-Zero – Automatic ZeroingThis function is used to tare small variations in weight automatically.In the event that small quantities are removed or added to the material to beweighed, incorrect weighing results can be displayed due to the “stabilitycompensation“ in the balance. (e.g. slow flow of liquids from a container placed onthe balance, evaporating processes).When apportioning involves small variations of weight, it is advisable to switch off thisfunction.Call up menu:Ö Press the + button simultaneously, "P1 rEAd" will appear⇒ Press repeatedly until the wantedsetting appears (no = off, yes = on)⇒⇒⇒⇒Ö Press the button repeatedly until the "SAVE"? inquiry appears.Confirm the inquiry by pressing the button to save the changes and toreturn to Weighing mode.-or-press the button to return to Weighing mode without saving.