PLE-BA-e-0812 5917 Instant helpIn case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.Help:Fault Possible causeThe displayed weight doesnot glow. • The balance is not switched on.• The mains supply connection has been interrupted(mains cable not plugged in/faulty).• Power supply interrupted.• (Rechargeable) batteries are inserted incorrectly orempty• No (rechargeable) batteries inserted.The displayed weight ispermanently changing • Draught/air movement• Table/floor vibrations• Weighing plate has contact with other objects.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)The weighing result isobviously incorrect • The display of the balance is not at zero• Adjustment is no longer correct.• Great fluctuations in temperature.• Warm-up time was ignored.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)Should other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.