PLE-BA-e-0812 1910 Basic Operation10.1 Switching on/offStart-up Press for approx. 0.5 sec.The balance will carry out a self-test The balance is ready forweighing when the weight display appears.Switching Off Press for approx. 0.5 sec.-OFF- appears briefly before the display disappears.10.1.1 Stability displayThe appearance of the stability symbol [ ] on the display indicates that theweighing plate is in a stable state. If the status is instable the [ ] displaydisappears.10.1.2 Balance zero displayShould the balance not display exactly zero despite empty balance pan, wait forstability display and press the button. The balance start with resetting to zeroand the symbol "a" will appear.Zeroing is only possible in the range ± 2% MAX (Err2).10.2 Simple weighing1. Place goods to be weighed on balance2. Wait until the stability display appears [ ]3. Read weighing result10.3 Weighing with taringThe dead weight of any weighing container may be tared away by pressing a buttonor by numerical input, so that the following weighings show the net weight of thegoods to be weighed.10.3.1 TaringÖ Put on weighing receptacles and press . The zero display and the symbolNet will appear.The tare weight is saved until it is deleted.Information:The tare procedure can be repeated as many times as necessary, for examplewith initial weighing of several components for a mix (add-on weighing). Thelimit is reached when the total weighing range capacity is full. After removingthe taring container the total weight is displayed as negative display.Taring is not possible with negative display values or zero display (Err3).