3Language Dictionary132Keysight Models 6811C, 6812C, and 6813C Programming GuideFUNCtion:TRIGgeredThis command selects the shape of the output voltage waveform when a triggeredstep or pulse transient occurs. The parameters are as follows:–SINusoid: A sine wave is the output.–SQUare: A square wave is the output.–CSINusoid: The output is a clipped sine wave. Both positive and negative peakamplitudes are clipped at a value determined by the FUNCtion:CSINusoidcommand.–
: The output shape is described by one of the user-defined waveformtables.The maximum peak voltage that the ac source can output is 425 Vpeak. Thisincludes any combination of voltage, voltage offset, and function shape values.Therefore, the maximum value that can be programmed depends on thepeak-to-rms ratio of the selected waveform. For a sine wave, the maximumvoltage that can be programmed is 300 Vrms.Command syntax[SOURce:]FUNCtion[:SHAPe]:TRIGgered ParametersSINusoid | SQUare | CSINusoid |