Using FunctionsSaving and Recalling a file45Saving and Recalling a fileN9330B allows you to save trace data, instrumentsetups, or the displayed screen graphic information intothe local memory of the tester. With a USB memory stickinserted, you can also save trace data and screen copiesinto an external memory devices.Trace data includes both the measurement data and theinstrument setup information. Trace data and instrumentsetup files are saved with a “.data” file extension.Save a trace dataTrace data includes the measured data and the currentinstrument setups. Save Data does not include limitlines. Limit lines can be saved using the Save Setupfeature.• Pressing [Save] > {Save Data} allows you to save thetrace data into local memory ONLY. It also activatesthe auto file naming editor and automatically savesthe current trace data with a name that is consecutiveto name of the latest saved file. This mode does NOT allow you to edit the file name.For example, if the previously saved file name is“Chicago_site1”, the current trace data will beautomatically named as “Chicago_site2”.• Pressing [Save] > {Save Data as} allows you to:• select a destination where a trace data is saved: Local or USB.• edit a user defined file name.See "Editing a file name" on page 47.File Format Max. # (Internal memory)Trace data *. data 200Instrument setting *.data 15Screen copy *.jpg Stored onto USB disk