Appendix C Common QuestionsVXI-MIO Series User ManualC-4 National Instruments Corporationa. Enable the PFI5 line for output, as follows:If you are using NI-DAQ, callSelect_Signal(deviceNumber, ND_PFI_5,ND_OUT_UPDATE, ND_HIGH_TO_LOW).If you are using LabVIEW, invoke the Route Signal VI withsignal name set to PFI5 and signal source set to AO Update.b. Set up data acquisition timing so that the timing signal forA/D conversion comes from PFI5, as follows:If you are using NI-DAQ, callSelect_Signal(deviceNumber, ND_IN_CONVERT,ND_PFI_5, ND_HIGH_TO_LOW).If you are using LabVIEW, invoke the AI Clock Config VIwith clock source code set to PFI pin, high to low, and clocksource string set to 5.c. Initiate analog input data acquisition, which will start onlywhen the analog output waveform generation starts.If you are using NI-DAQ, call DAQ_Start with appropriateparameters.If you are using LabVIEW, invoke the AI Control VI withcontrol code set to 0 (start).d. Initiate analog output waveform generation.If you are using NI-DAQ, call WFM_Group_Control withoperation set to 1 (start).If you are using LabVIEW, invoke the AO Control VI withcontrol code set to 0 (start).Timing and Digital I/O12. What types of triggering can be hardware implemented on myVXI-MIO Series module?Digital triggering is supported by hardware on everyVXI-MIO Series module. In addition, the VXI-MIO-64E-1 andVXI-MIO-64XE-10 support analog triggering in hardware.13. What added functionality does the DAQ-STC make possible incontrast to the Am9513?The DAQ-STC incorporates much more than just 10 Am9513-stylecounters within one chip. In fact, the DAQ-STC has the complexityof more than 24 chips. The DAQ-STC makes possible PFI lines,