EN-19Adventurer Pro Grain Test Scale3.7.3 SetupLegal TradeWhen Comm is selected, the scale is suitablefor commercial grain inspection. When GIPSAis selected, the scale is suitable for officialGIPSA grain inspection.Auto OffWhen set to 1, 2 or 5 minutes, the scale willturn off in the selected time interval if there hasbeen no activity. Set to OFF for continuousoperation.End SetupPress the Yes button, to advance to the Readoutsubmenu. Press the No button to return to theLEGAL TRADE menu item.Auto ZeroAllows setting the scale auto zero level setting:.5, 1, 2 or 5 divisions. Scale maintains zerountil threshold is exceeded.FilterAllows setting the scale filter level: Low,Medium, High. Compensates for vibration orexcessive air currents.READOUTAuto Zero Off, .5d, 1d, 2d, 5dFilter Low, Medium, HighEnd Read3.7.4 Readout3.7.5 ModeThe Mode submenu is used to set Applicationmodes parameters.MODEWeigh On/OffTest Weight lb/bu (qt)lb/bu (pt)kg/hL (L)kg/hL (.5L)Dockage Pos, NegEnd ModeWeighUse to set the Weighing Mode On or Off.Test WeightUse to set the Test Weight to: lb/bu (qt),lb/bu (pt),kg/hL (L) or kg/hL (.5L)DockageUse to set Dockage Mode to: Positive ornegative.End ModePress the Yes button to advance to the Unitsubmenu. Press No to return to Weigh.3.7.6 UnitThe Unit submenu is used to turn measuringunits On or Off.End ReadPress the Yes button to advance to the Modesubmenu. Press No to return to Auto Zero.UNITg On/Offkg On/Offoz On/Offlb On/OffGN On/OffCustom On/OffEnd UnitSETUPLegal Trade Off, Comm,GIPSAAuto Off Off, 1,2,5End Setup