21c) If you have selected multiple contrast methods that require distinctillumination methods (e.g. Bright Field and Fluorescence), you must use the Wait block.To set a Wait block, just click in the Wait button.As it is strictly required to operate the diaphragm slide between transitions oftransmitted and incident light (e.g. from bright field to fluorescence GFP), the Wait blockwill make a pause during the experiment.At the pause, you should set the diaphragm slide in agreement with the nextcontrast method.To switch the order among blocks, just click in a block and move it into thedesired position of the timeline of events.10. Box the 1 separate image document / acquisition block option to create animage for each block of the timeline of events